Cryo-EM Dose Rate Calculator
The SubAngstrom® Electron Flux Calculator works out the electron dose of a cryo-EM exposure. Enter the dose rate on the detector, the pixel size, the exposure time and the frame time, and it gives you the dose per Ų per second, the dose per frame, the total dose and the number of frames.
The calculator is free with an approved SubAngstrom account, and accounts are usually approved the same day. Logged in, you can also email the results to yourself or your team, or download them as an image.
Log in to use the calculator.
What you enter
- Dose (electrons/pixel/second): the dose rate on the detector
- Pixel size (Å/pixel)
- Exposure time (seconds)
- Frame time (seconds)
- Tilt range ± (degrees): tomography only
- Tilt increment (degrees): tomography only
How it is worked out
- Pixel size squared = pixel size × pixel size
- Dose per second = dose ÷ pixel size squared
- Dose per frame = dose per second × frame time
- Total frames = exposure time ÷ frame time
- Total dose = dose per second × exposure time
Tick Tomography to add a tilt range and tilt increment. The calculator then counts the tilts (2 × tilt range ÷ tilt increment + 1), gives the dose per tilt, and multiplies the dose per tilt by the number of tilts for the total dose of the tilt series.
Worked example
A dose of 16 electrons/pixel/second at a pixel size of 0.8 Å/pixel, with a 2 second exposure split into 0.05 second frames:
| Pixel size squared | 0.8 × 0.8 = 0.64 Ų |
|---|---|
| Dose per second | 16 ÷ 0.64 = 25 electrons/Ų/second |
| Dose per frame | 25 × 0.05 = 1.25 electrons/Ų/frame |
| Total frames | 2 ÷ 0.05 = 40 frames |
| Total dose | 25 × 2 = 50 electrons/Ų |